Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Sub Wait for 792A to Settle /8506 DATE: 02-Oct-98 AUTHOR: Fluke REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 64 CONFIGURATION: Fluke 8506A ============================================================================= # # Source: # # Compatibility: # MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # None # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 MATH M[17] = 0 1.002 8506 0V N 2W 1.003 MATH M[18] = MEM 1.004 8506 0V N 2W # Determine frequency range 1.005 MATH MEM1 = M[2] 1.006 JMPZ 1.022 1.007 MATH MEM1 = M[2] - 20E+3 1.008 JMPT 1.022 1.009 MATH MEM1 = M[2] - 90 1.010 JMPT 1.019 1.011 MATH MEM1 = M[2] - 15 1.012 JMPT 1.016 # <=15Hz, consecutive readings MEM and M[18] must be within 4uV 1.013 MATH MEM1 = MEM - M[18] 1.014 JMPZ 1.029 4e-006U 1.015 JMP 1.024 # >15Hz & <=90Hz, consecutive readings MEM and M[18] must be within 2uV 1.016 MATH MEM1 = MEM - M[18] 1.017 JMPZ 1.029 2e-006U 1.018 JMP 1.024 # >90Hz & <=20kHz, consecutive readings MEM and M[18] must be within 1uV 1.019 MATH MEM1 = MEM - M[18] 1.020 JMPZ 1.029 1e-006U 1.021 JMP 1.024 # >20kHz, consecutive readings MEM and M[18] must be within 2uV 1.022 MATH MEM1 = MEM - M[18] 1.023 JMPZ 1.029 2e-006U 1.024 MATH M[17] = M[17] + 1 1.025 MATH MEM1 = M[17] - 10 1.026 JMPF 1.003 1.027 OPBR Ten readings have been taken and the 792A has not 1.027 OPBR settled to within the specified limits. 1.027 OPBR 1.027 OPBR Do you wish to PROCEED anyway? 1.028 JMPF 1.001 1.029 END